What need to check before starting the heater and chiller?

views:381

  • Heating and cooling circulator is widely matched for low temperature ethanol extraction equipment, rising film evaporator, wiped film evaporator, rotary evaporator, decarboxylation reactor, short path distillation equipment, crystallization reactor, glass reactor, stainless steel reactor, etc. To achieve precise temperature control.

     


  • What need to check before starting the heater and chiller?


    1.Please check whether the outer packaging is intact, and check whether the electrical equipment is loose, unloading or falling off, if any, please contact TOPTION in time info@toptionlab.com 


    2. Remove the inlet & outlet of the heater and chiller, including the oil drain plug (Please note: There is a little heat transfer oil left in machine during machine testing in factory, please pay attention to prevent dripping).



    3. Connect the power supply, then open the control box and check whether the phase sequence indicator light is on, If it is on, it means that the power supply wiring is correct before starting the heater / chiller.



    4. After the above inspection is completed, you can start to add heat transfer oil or circulating fluid, the filling port is at the upper part, first open the exhaust valve, instantaneous needle rotation is closed, counterclockwise rotation is open.


    5. Pay attention to the liquid level indicator when adding liquid. When the liquid level indicator is at 2/3, the machine turns on the circulation button on the panel, starts the circulating pump, and the liquid level begins to drop. At this time, pay attention to refilling. When the circulation is full, the liquid level indicator Should be at 1/2 of the fuel tank. 


    6. The first use keeps the cycle time not less than 15 minutes, and the heating or cooling can be started after 15 minutes.


    8. Pay attention to the first heating, when the temperature is above 80℃, the heating setting temperature range should be small, It is recommended to circulate at 80°C-110°C for half an hour, 110°C to 130°C for half an hour, and 130°C to 150°C for half an hour. ( This is to prevent too much moisture in the heat transfer oil and affect the equipment cycle. Not needed for cooling).


    9.After the above operations are completed, close the top exhaust valve and start normal use.

     

  • TOPTION as manufacturer of extraction | distillation equipment and heating & cooling circulators , with 15+ years development, production experience, provide technical support for new starter, professional customize, wholehearted service, contact info@toptionlab.com  to get your one stop solution design. 

       



Comment



Popular Blog

  • > Chemical Sample Pre-treatment Technology Analysis From TOPTION Specialist
  • > Chemical Jacketed Lab Reactors,Downward Open Reactor Customize
  • > How Choose Diaphragm Metering Pump For Chemical Glass Reactor?
  • > How To Use Quartz Photochemical Reactor ?
  • > Glass Vessel Chemical Reactors , Glass Reactor Suppliers In China
  • > How To Operate Chemical Glass Reactor? Glass Reaction Vessel Operating
  • > 800L Jacketed Glass Reactor Design, Where Customize Glass Chemical Reactor?
  • > 1L Double Layer Glass Reactor New Customize, Lab Chemical Reaction Vessel
  • > Photocatalytic reactor setup photochemical reactor suppliers


  • > Vacuum Pump Maintenance And Repair Manual From Pumping Machine Manufacturer
  • > Membrane Vacuum Pump Diaphragm Vacuum Pump Manufacturers
  • > Low Temperature Cooling Liquid Circulating Pump Manufacturer China
  • > Laboratory Thermostatic Closeness Ultrasonic Reactor Manufacturer
  • > How Much 500w Xenon Lamp Price? Xenon Light Source Manufacturer
  • > Small Spray Dryer, Vegetable Extracts Spray Dryer Manufacturer
  • > How To Clean Vacuum Filter Vacuum Filtration Apparatus Manufacturer
  • > Fermentor For Sale Fermentor Tank Manufacturer
  • > High Shear Homogeneous Reactor System Manufacturer


  • > How Choose Gas Chromatography Machine? TOPTION Lab Instrument Supplier
  • > How To Use Thermostatic Shaker Safety? TOPTION Lab Shaker Supplier
  • > How To Use High Pressure Homogenizer, Lab Homogenizer Supplier
  • > High Shear Mixer Supplier In China, Find High Shear Mixers Distribute
  • > What's The Ultrasonic Cell Crusher Cost? Ultrasonic Liquid Processors Supplier
  • > What Can Ultrasonic Homogenizer Do? Ultrasonic Homogenizer Supplier
  • > How To Use Ultrasonic Cleaning Machine? Ultrasonic Cleaner Supplier
  • > How To Fit Glass Ultrasonic Horn? Ultrasonic Homogenizer Supplier
  • > Short Time Mini Dryer Milk Powder Spray Dryer Supplier
  • Share This Solution To Your Advantages.

    CONTACT US

    Email NewsLetter

    Your info are absolutely safe.

    Free&Professional report of chinese lab instruments market.

    Copyright 2022 © TOPTION INSTRUMENT CO.,LTD. All Right Resrrved  | Privacy Policy

    Send Inquiry

    Send your inquiry

    X